MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number S-60-100-120-P6
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Improper or Incorrect Procedure or Method (2017); Difficult or Delayed Activation (2577)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 10/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during deployment the stent inadvertently partially deployed in the sheath resulting in the reported difficult or delayed activation and the reported difficult to remove.Interaction of devices/manipulation using force resulted in the reported tip material separation.The treatment(s) appears to be related to the operational context of the procedure as reportedly two snares were attempted and other balloon techniques, but the tip could not be removed.The tip was able to be dragged to the left common femoral via the left groin access and the tip was embedded to the vessel with an additional stent.There was a 2+ hour delay in the procedure but this did not cause harm to the patient.Reportedly, the physician pulled hard when some resistance was noted during removal and the tip separated.It should be noted that the supera peripheral stent system instructions for use states: should unusual resistance be felt at any time during stent system removal, the entire system should be removed together with the introducer sheath or guiding catheter as a single unit.Although it was noted the physician pulled hard in the attempt to remove the device the force applied seemed to be a reasonable clinical response to the difficulties.There is no indication of a product quality issue with respect to manufacture, design or labeling.(b)(4).
 
Event or Problem Description
It was reported the procedure was to treat the right superficial femoral artery (sfa) via left groin access with calcification and tortuosity noted in the vessel.The vessel was 6 mm, atherectomy was not used and the vessel was prepared with a 6x200 mm armada 18 percutaneous transluminal angioplasty (pta) catheter to 14 atmospheres at 3 minutes.The 6x100 mm supera self expanding stent (ses) was thought to have been successfully deployed; however, it partially deployed in the 6fr 45 cm sheath and upon removal the tip of the delivery system caught on the sheath.The physician pulled hard when some resistance was noted during removal and the tip separated.Two snares were attempted and other balloon techniques but the tip could not be removed.The tip was able to be dragged to the left common femoral via the left groin access and the tip was embedded to the vessel with an additional stent.There was a 2+ hour delay in the procedure as a result of the additional intervention.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15633843
Report Number2024168-2022-10822
Device Sequence Number18144425
Product Code NIP
UDI-Device Identifier08717648211829
UDI-Public08717648211829
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberS-60-100-120-P6
Device Catalogue NumberS-60-100-120-P6
Device Lot Number2081761
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2022
Initial Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6FR 45 CM DESTINATION SHEATH
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
Patient Weight93 KG
-
-