Model Number 15624K
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Failure of Implant (1924); Inadequate Osseointegration (2646)
Date of Event 09/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Failed to osseointegrate.
 
Additional Manufacturer Narrative
Keystone dental inc.Provides labeling with all implants.Failure to osseointegrate and loss of osseointegration (implant mobility) is identified as an adverse reaction and identified in the list of warnings in all endosseous dental implant labeling.The specific cause for a particular complaint is often not readily identified as there are various factors that contribute to the risk of implant failure.These include patient factors such as prior oral infection, poor bone quality or quantity, systemic conditions such as diabetes and uncontrolled hypertension.Technique factors such as overheating of the implant site may cause necrosis of the bone, preventing growth (pelayo et al., 2008; turkyilmaz & al., 2008).Patient habits such as tobacco use (heitz-mayfield & huynh-ba, 2009), alcohol or drug abuse, poor oral hygiene, and bruxism (salvi & braegger, 2009) may also lead to implant failure.In addition, improper surgical technique can lead to implant failure and/or loss of supporting bone (salvi & braegger, 2009).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMA
Common Device Name
PRIMACONNEX TC TAPERED WD 5.0X15
Manufacturer (Section D)
KEYSTONE DENTAL, INC.
154 middlesex turnpike
burlington MA 01803
Manufacturer (Section G)
KEYSTONE DENTAL, INC.
154 middlesex turnpike
burlington MA 01803
Manufacturer Contact
ryan raymond
154 middlesex turnpike
burlington, MA 01803
MDR Report Key15633867
Report Number3005990499-2022-14267
Device Sequence Number15446286
Product Code DZE
UDI-Device IdentifierD76815624K0
UDI-Public+D76815624K0/$$324053144827G
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized123
Summary Report (Y/N)Y
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Dentist
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/30/2024
Device Model Number15624K
Device Catalogue Number15624K
Device Lot Number44827
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2022
Initial Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/07/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient SexMale
-
-