SYNTHES GMBH RFNA/ 9MM / 160MM/ STANDARD BEND/ STER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
04.233.916S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: only the event year is known.Complainant part is not expected to be returned for manufacturer review/investigation.Additional device product codes: hwc.Manufacturing location: monument.Manufacturing date: 16-sep-2021.Expiration date: unknown.Part number: 04.233.916s, rfn/9mm/160mm 5 degree bend/pe inlay/sterile.Lot number: 288p783 (sterile).Lot quantity: 6.Production order traveler met all inspection acceptance criteria.Inspection sheet, in-process / inspect dimensional / final ns499737 rev c met all inspection acceptance criteria.Inspection sheet, rfn assembly inspection, ns529720 rev b met all inspection acceptance criteria.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Note: there was no scn number recorded on the production order traveler.Therefore, the scn for this lot could not be reviewed.There was no pll included in this dhr.Therefore, the pll for this lot could not be reviewed and the expiration date could not be notated.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: part number: 04.233.124.2y, poly inlay retrograde femoral.Lot number: 182p313.Lot quantity: 120.Production order traveler met all inspection acceptance criteria.Inspection sheet, ns074000 rev a met all inspection acceptance criteria.Inspection sheet, ns074022 rev a met all inspection acceptance criteria.Part number: 21131, tialv*ri13.00.Lot number: 64p4762.Lot quantity: 2,341 lbs.Inspection instruction met all inspection acceptance criteria.Product certification supplied by nf&m international inc.Dated 01-jun-2020 was reviewed and determined to be conforming.Lot summary report dated 28-jul-2020 met all inspection acceptance criteria.Raw material receiving / putaway checklist met all inspection acceptance criteria.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, the physician reported that a short rfna which he had implanted on (b)(6) 2022 had broken at the junction between the most distal screw of the proximal interlocking screws.The surgeon removed the broken hardware on (b)(6) 2022 and revised the construct with a nail and plate.No further information is available.This report involves one rfna / 9mm / 160mm 5 degree bend / sterile.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: h3, h4, h6: a manufacturing record evaluation was performed: product code: 04.233.916s, lot: 288p783: manufacturing location: monument.Manufacturing date: september 16, 2021.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.H3, h6: a product investigation was completed: visual analysis of the returned sample revealed that the rfna was found broken from the shaft.The broken fragment was returned for evaluation.No other defect was found.A dimensional inspection was not performed as it is not applicable to the complaint condition.The current and manufactured to drawing was reviewed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|