| Model Number |
GMM-344L |
| Medical Device Problem Codes |
Break (1069); Product Quality Problem (1506)
|
| Health Effect - Clinical Codes |
Necrosis (1971); Hernia (2240); Peritonitis (2252); Unspecified Tissue Injury (4559)
|
| Date of Event |
08/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, post operatively, on an exploratory laparotomy, while on fascial closing, the doctor used two looped sutures to close the fascia; one at each end of the incision.The stitches were started with a v stitch then closed by the resident in a standard continuous fashion.The surgeon tied the knots at the end.The two threads broke.The sutures were binding on itself and felt stickier than usual.After the user washed and closed the wound, it was left open to prevent infection.No steroids have been given to the patient and was still admitted.There was an unanticipated tissue loss and tissue damage.There was dead fascia which left the patient with a ventral hernia.It was noted there will be an additional operation performed to correct the issue.It was noted that the patient was critically ill.
|
| |
|
Event or Problem Description
|
|
According to the reporter, post operatively, on an exploratory laparotomy, while on fascial closing, the doctor used two looped sutures to close the fascia; one at each end of the incision.The stitches were started with a v stitch then closed by the resident in a standard continuous fashion.The surgeon tied the knots at the end.The two threads broke.The sutures were binding on itself and felt stickier than usual.After the user washed and closed the wound, it was left open to prevent infection.No steroids have been given to the patient and was still admitted.There was an unanticipated tissue loss and tissue damage.There was dead fascia which left the patient with a ventral hernia.It was noted there will be an additional operation performed to correct the issue.It was noted that the patient was critically ill with chronic peritonitis.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the suture thread broke and the suture did not appear as expected.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, post operatively, on an exploratory laparotomy, while on fascial closing, the doctor used two looped sutures to close the fascia; one at each end of the incision.The stitches were started with a v stitch then closed by the resident in a standard continuous fashion.The surgeon tied the knots at the end.The threads broke.The sutures were binding on itself and felt stickier than usual.After the user washed and closed the wound, it was left open to prevent infection.No steroids have been given to the patient and was still admitted.There was an unanticipated tissue loss and tissue damage.There was dead fascia which left the patient with a ventral hernia.It was noted there will be an additional operation performed to correct the issue.It was noted that the patient was critically ill with chronic peritonitis.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: gmm-344l, gmm-344l maxon* 1 grn 240cm lp gs26x12, (lot#d2f2443y); gmm-344l, gmm-344l maxon* 1 grn 240cm lp gs26x12, (lot#d1m2128y).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|