|
Product complaint # (b)(4).No device problem found.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that it was received one empty open folder and one needle-suture piece that pertain to product code sxpp1a415.During the visual inspection of the two dispensed needle-suture combinations, the swage and attachment area were noted as expected.The sutures were examined, and the barbs were present along of the strand; also, 10 barbs were measured, and all barbs met with the requirements.In addition, the loops were as expected.As part of quality process, all devices are manufactured, inspected, and released to approved specifications.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 2360 g/m.Events reported via: 2210968-2022-08562.
|