MAUDE Adverse Event Report: ETHICON INC. STFX SYM PDS+ UNI VIO 6IN 2-0 S/A CT-2; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. STFX SYM PDS+ UNI VIO 6IN 2-0 S/A CT-2; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE Back to Search Results
Model Number SXPP1A415
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a gastric bypass procedure in 2022 and barbed suture was used.During the procedure, it was reported that the barbs are pulling right through and not catching in the tissue.A vloc was used to close.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).No device problem found.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that it was received one empty open folder and one needle-suture piece that pertain to product code sxpp1a415.During the visual inspection of the two dispensed needle-suture combinations, the swage and attachment area were noted as expected.The sutures were examined, and the barbs were present along of the strand; also, 10 barbs were measured, and all barbs met with the requirements.In addition, the loops were as expected.As part of quality process, all devices are manufactured, inspected, and released to approved specifications.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 2360 g/m.Events reported via: 2210968-2022-08562.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STFX SYM PDS+ UNI VIO 6IN 2-0 S/A CT-2
Common Device Name
SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC SAN LORENZO
road 183, km. 8.3
san lorenzo PR
*  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15633892
Report Number2210968-2022-08561
Device Sequence Number7970620
Product Code NEW
UDI-Device Identifier10705031244986
UDI-Public10705031244986
Combination Product (Y/N)Y
Initial Reporter CountrySF
PMA/510(K) Number
K141776
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model NumberSXPP1A415
Device Catalogue NumberSXPP1A415
Device Lot NumberRLBDLX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Initial Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-