| Model Number |
MF-410255 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported right side capsular contracture baker grade unknown.Patient additionally reported "firmness, and hardness", the device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown.
|
| |
|
Event or Problem Description
|
|
Healthcare professional later reported capsular contracture baker grade iv and exchange from textured to smooth due to concern with the product.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, g2.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced with a non-allergan device.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, h6.
|
| |
|
Event or Problem Description
|
|
Patient reported right side capsular contracture baker grade unknown.Patient additionally reported "firmness, and hardness".Healthcare professional later reported capsular contracture baker grade iv and exchange from textured to smooth due to concern with the product.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: anxiety product/procedure: unable to observe as it is not related to the device.Capsular contracture: unable to observe as it is not related to the device.Additional observations.Observed 1 opening assessed as fold flow opening.Stress marks observed are assessed as non-penetrating nick.No further actions are required as no manufacturing issues are observed.
|
| |
|