| Model Number |
ILPC442U |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).Patient id not provided.Patient weight not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the abutment screw was found loose at the site.Upon investigation it was actually fractured and was removed and replaced.
|
| |
|
Event or Problem Description
|
|
Per investigation results the abutment is not fractured but the retaining screw.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).Per investigation findings the abutment reported as a fractured is not fractured.An additional retaining screw was found fractured.This event is therefore no longer reportable , no additional reports would be submitted.
|
| |
|