Model Number TPL0059
Medical Device Problem Codes Imprecision (1307); Unintended Collision (1429)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A medtronic representative went to the site to test the equipment.At the time of the system checkout the surgical arm was replaced.The system then passed the system checkout and was found to be fully functional.Concomitant medical products: other relevant device(s) are: pn: asm0206, sn: unk.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
Medtronic received information regarding a guidance system being used during a spinal procedure.It was reported that while in a case the surgeon was performing a open procedure l3-s1.Placing competitive guidewires and all the wires were determined to be off (10 wires).The mnaufacturer representative thought roughly 3-5mm off.L2, l3, and l4 were all low.L5 and s1 medial on right, lateral on left.The inaccuracies were inconsistent between operating levels.The surgeon took c-arm shots after placing guidewires and that is when it was determined that all wires were inaccurate.At this point they pulled the wires and went to flouro. during the case manufacturer representative stated when they sent the arm, it hit trajectory, he used the down arrow to lower the arm closer to patient.He did that to bring trajectory closer.As he watched some of those motions from the side he thinks you can see the arm shake just a little bit.Enough for him to catch it (no canula in arm guide). after the case, the manufacturer representative put the system in demo mode, sent the arm to trajectory, dropped pointer (from door bist test) in guide to mark where it hit the table.He then pulled out, lowered arm close down, put pointer back in and it had moved off trajectory slightly.The issues extended the surgical time by less than 1 hour.There was no impact on the patient outcome.
 
Additional Manufacturer Narrative
H2) correction: d9, return date corrected.The part is currently under analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
H3: the surgical arm was returned for analysis and multiple hardware failures were confirmed with the part.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: pn: asm0206, sn: (b)(6).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAZOR X SYSTEM
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IS  3079567
Manufacturer (Section G)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IS   3079567
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15634435
Report Number3005075696-2022-00074
Device Sequence Number20825811
Product Code OLO
UDI-Device Identifier07290109184098
UDI-Public07290109184098
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K182077
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTPL0059
Device Catalogue NumberTPL0059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2022
Supplement Date Received by Manufacturer11/17/2022
12/15/2022
Initial Report FDA Received Date10/19/2022
Supplement Report FDA Received Date12/07/2022
01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
Patient Weight94 KG
-
-