MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ENDOTRACHEAL TUBES SILICONIZED; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ENDOTRACHEAL TUBES SILICONIZED; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 100/166/080
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No lot number was provided; therefore, device history record review could not be completed.One sample returned in used condition and without its original packaging.Visual inspection performed at a distance of 12 to 16 inches and normal conditions of illumination and no discrepancies were observed.Functional testing of the returned sample was inflated using a syringe in order to detect any leakage.And there leakage was observed coming out from the seal of the cuff.The root cause was linked to manufacturing.Actions were taken to mitigate the reported issue.Per previous in-process discrepancy report update the procedure was updated and re-training of manufacturing personnel occurred.This mdr was generated under protocol b10009704, as a result of warning letter cms# 617147.
 
Event or Problem Description
It was reported that during a pre-use check on the device, the customer noticed the cuff would not inflate.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ENDOTRACHEAL TUBES SILICONIZED
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15635245
Report Number3012307300-2022-25206
Device Sequence Number12211775
Product Code BTR
UDI-Device Identifier15019315054105
UDI-Public15019315054105
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K083119
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100/166/080
Device Catalogue Number100/166/080
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2022
Initial Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-