Model Number 364902
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd luer-lok¿ access device (male luer), there was blood leakage or other sample leakage from the device other than the insertion site or needle tip.This event occurred 2 times.The following information was provided by the initial reporter.The customer stated: "whilst collecting a patient sample, the llad was leaking from the 18g needle inside the holder where the blood tubes are inserted.Customer collected 2-3 blood tubes and not sure whether this is caused by user error or the llad is leaking.She did note this has happened in 2 separate patients and could be a device error.
 
Additional Manufacturer Narrative
H.6.Investigation summary: bd had not received samples, but 3 photos were provided for investigation.The photos were reviewed and the indicated failure mode for leakage was observed.Additionally, 48 retention samples from bd inventory were evaluated by visual examination and 12 retention samples by functional testing and the issue of leakage was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode leakage.Bd was not able to identify a root cause for the indicated failure mode.H3 other text : see h.10.
 
Event or Problem Description
It was reported when using the bd luer-lok¿ access device (male luer) there was blood leakage or other sample leakage from the device other than the insertion site or needle tip.This event occurred 2 times.The following information was provided by the initial reporter.The customer stated: "whilst collecting a patient sample the llad was leaking from the 18g needle inside the holder where the blood tubes are inserted.Customer collected 2-3 blood tubes and not sure whether this is caused by user error or the llad is leaking.She did note this has happened in 2 separate patients and could be a device error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD LUER-LOK¿ ACCESS DEVICE (MALE LUER)
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
Manufacturer (Section G)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15635251
Report Number2618282-2022-00068
Device Sequence Number12391753
Product Code JKA
UDI-Device Identifier50382903649025
UDI-Public50382903649025
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K991088
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2024
Device Model Number364902
Device Catalogue Number36490200
Device Lot Number1174286
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2022
Supplement Date Received by Manufacturer10/20/2022
Initial Report FDA Received Date10/19/2022
Supplement Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-