| Model Number |
0502537030 |
| Medical Device Problem Code |
Fogging (1253)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was foggy image during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for investigation and the reported failure mode was not confirmed.Alleged failure: fogging.Probable root cause: laser welding seal failure; distal/proximal window solder failure; damage to optical train; damage to needle; damage to distal or proximal windows; moisture intrusion; end of life wear-out; environmental disturbance: endoscope colder than dew point; environmental disturbance: fluid entrapment between the coupler and eyepiece junction (eyepiece only); use error.The device manufacture date is not known.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
|
| |
|