• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. D-RAD SMART PACK 4H LEFT STANDARD PLATE; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. D-RAD SMART PACK 4H LEFT STANDARD PLATE; PLATE, FIXATION, BONE Back to Search Results
Model Number 71158019
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Rupture (2208)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, following an internal fixation surgery performed on an unspecified date to treat a radius fracture with a d-rad single use kit plating system, the patient experienced the rupture of the flexor pollicis longus tendon which required the plate removal on an unknown date.The patient had achieved radiographic union of the fracture 1.5 months post initial implantation.No further information is available.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, as of the date of this investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The data was collected from the post-market clinical follow up activity (pmcf) was anonymous, no further information will be provided.Consequently, without the requested clinical information, a thorough clinical assessment cannot be performed at this time.The impact to the patient beyond that which has already been reported cannot be confirmed nor concluded based on the information provided.Therefore, no further medical assessment is warranted at this time.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for d-rad smart pack¿ system provides complete guidelines of indications, contraindications, warnings and precautions and possible adverse effects that may occur preoperative, during surgery or post operative.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, patient condition or postoperative care.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
 
Additional Manufacturer Narrative
Corrected data: b5 (narrative), g2 (source: pmcf conducted in the uk), h6 (health effect - impact code).H11: additional information has been received from the site conducting this pmcf.
 
Event or Problem Description
It was reported that, following an internal fixation surgery performed on an unspecified date to treat a radius fracture with a d-rad single use kit plating system, the patient experienced the rupture of the flexor pollicis longus tendon at three (3) months postoperatively which required the plate removal and thumb ipj fusion an unknown date at six (6) months postoperatively.The patient had achieved radiographic union of the fracture 1.5 months post initial implantation.The patient's symptoms have completely resolved and is back to usual activities.No further information is available.
 
Additional Manufacturer Narrative
Corrected data: h6 (health effect - impact code).Updated results of investigation: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, there were no clinical factors found which would have caused or contributed to the reported fracture collapse.Consequently, a thorough medical investigation cannot be rendered, nor could a definitive clinical root cause of the reported failure be determined.Per the report, the patient experienced the rupture of the flexor pollicis longus tendon at three (3) months postoperatively which required the plate removal and thumb interphalangeal joint fusion an unknown date at six (6) months postoperatively.It was also reported the patient achieved union of the fracture.Therefore, impact to the patient beyond that which has already been reported cannot be confirmed nor concluded no further clinical/medical assessment is warranted at this time.Should any additional relevant medical information be provided, this case would be re-assessed.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for d-rad smart pack¿ system provides complete guidelines of indications, contraindications, warnings and precautions and possible adverse effects that may occur preoperative, during surgery or post operative.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, patient condition or postoperative care.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
D-RAD SMART PACK 4H LEFT STANDARD PLATE
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15636028
Report Number1020279-2022-04500
Device Sequence Number7179113
Product Code HRS
UDI-Device Identifier00885556364444
UDI-Public00885556364444
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K132296
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71158019
Device Catalogue Number71158019
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer10/21/2022
03/20/2023
05/18/2023
Initial Report FDA Received Date10/19/2022
Supplement Report FDA Received Date10/26/2022
03/31/2023
05/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-