| Model Number |
133S-MV-11-T |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: foreign material on implant.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported opening a "sterile in the box implant stating "thought it was used", ¿there are marks on the implant.Must be a manufacturer mistake." device was never implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: foreign material on implant : material of the stable base is observed adhered to the internal part of the shell additional observations: no device failure observed.Additional, changed, and/or corrected data: d9, h3, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported opening a "sterile in the box implant" stating "thought it was used, there are marks on the implant.Must be a manufacturer mistake." device was never implanted.
|
| |
|