| Model Number |
0502123030 |
| Medical Device Problem Code |
Moisture or Humidity Problem (2986)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|
Additional Manufacturer Narrative
|
|
This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke service record attached, the reported failure "condensation" was confirmed.According to henke: scope evaluated as a level 3.Bent, dent, tip and fiber damaged, particulates under distal lens and proximal lens, broken lenses in system, scratches, residue on distal lens negative, moisture in system and moisture under distal lens negative.Probably root cause for this failure is: the complaint for ¿eib marjorie materials mgnmt, has condensation¿ has been confirmed and is due to customer use and handling.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|