| Model Number |
E601 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The calibration and qc results were within the acceptable range.Therefore, a general reagent issue could be excluded.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|
Event or Problem Description
|
|
There was an allegation of a questionable ca 19-9 elecsys result from the cobas 6000 e601 module.The initial result was 40.32 u/ml which was considered wrong and not released to the patient or doctor.The repeat result from another cobas e601 analyzer was 9.89 u/ml.The repeat result on the same analyzer was 9.12 u/ml with a data flag and then 9.13 u/ml.This result was considered correct and released to the patient.The reagent lot number was 598103 with an expiration date of 31-aug-2023.
|
| |
|