DENTSPLY IMPLANTS MANUFACTURING GMBH UNKNOWN ATIS EV ABUTMENT; ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS Back to Search Results
Catalog Number UNK ATIS EV ABUTMENT
Device Problem Fracture (1260)
Patient Problems Failure of Implant (1924); Unspecified Infection (1930); Inflammation (1932); Pocket Erosion (2013)
Event Date 09/06/2022
Event Type  Injury  
Manufacturer Narrative
Therefore, because a serious injury resulted, this event is reportable per 21 cfr part 803.Section was done based on the information provided by the initial reporter and our long-time experience in the investigation of similar complaints.Product return is requested and product will be evaluated after receipt.In case any new or additional information will be gained from this investigation a follow-up report will be sent.Trend is tracked and monitored.
 
Event Description
It was reported that a patient experienced a dental implant loss.
 
Manufacturer Narrative
The optical investigation showed that the implant is not fractured but was removed due to the fracture of a screw or one-piece abutment.Corrected product information, lot number and concomitant product in this report.This report is to also correct the following codes: component code: add 568; type of investigation: add 10 and 4116; investigation findings: remove 3243.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN ATIS EV ABUTMENT
Type of Device
ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS
Manufacturer (Section D)
DENTSPLY IMPLANTS MANUFACTURING GMBH
rodenbacher chaussee 4
hanau hessen 63457
GM  63457
Manufacturer (Section G)
DENTSPLY IMPLANTS MANUFACTURING GMBH
rodenbacher chaussee 4
hanau hessen 63457
GM   63457
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key15637971
MDR Text Key302057133
Report Number3013111692-2022-21692
Device Sequence Number1
Product Code NHA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120414
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue NumberUNK ATIS EV ABUTMENT
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date10/17/2022
Date Manufacturer Received10/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
Treatment
OSSEOSPEED EV 4.8 S - 9 MM
Patient Outcome(s) Required Intervention;
Patient Age40 YR
Patient SexFemale
-
-