• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE Back to Search Results
Model Number 777424
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the bracket of the ureteral stent had high friction and could not be put in.The user changed to a new device.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned and further evaluation was not conclusive.Although a specific cause cannot be determined, based on the risk document a potential root cause for this event could be, "material selection".The device was attempted to be used for treatment purposes.It is unknown if the device had met all relevant specifications or resulted in the reported event.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "the insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure." "directions for use: 1.Determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram.Accurate measurements will optimize drainage efficiency and patient comfort.Submerge stent in sterile water to activate the coating.2.Insert the cystoscope then pass the guidewire* through the scope until the tip is in the renal pelvis.3.Move the pigtail straightener over the proximal end (kidney coil end) of the ureteral stent allowing easier insertion onto the guidewire.Remove pigtail straightener once the stent is secure on the guidewire.4.Pass the stent over the guidewire through the cystoscope by using the push catheter for proper placement.5.Watch the distal end (bladder coil end) of the stent or the radiopaque, proximal end of the pusher.Stop advancing when the stent¿s distal end marker reaches the ureterovesical junction (uvj).6.Withdraw the guidewire slowly.The stent will form a pigtail automatically.7.Carefully remove the push catheter." "*activate the guidewire coating according to the ¿instructions for use¿ found within the guidewire packaging." the device was not returned.
 
Event or Problem Description
It was reported that the bracket of the ureteral stent had high friction and could not be put in.The user changed to a new device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15638213
Report Number1018233-2022-08013
Device Sequence Number6417713
Product Code FAD
UDI-Device Identifier10801741014472
UDI-Public(01)10801741014472
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K983498
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2026
Device Model Number777424
Device Catalogue Number777424
Device Lot NumberNGGR3325
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer02/27/2023
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date02/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-