| Model Number |
H-1129 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device had a sticky needle on the pressure gauge.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No causes or potential causes to the customer's reported problem were found during the device history record (dhr) review.The device was received for evaluation; in good condition.Device was manufactured in 2019.Visually inspected device and found that hi-2 cuffs pressure gauge needle sticking.This confirmed the reported issue.The root cause of the issue was deemed to be supplier related.Action(s) taken to mitigate the reported issue(s): replaced hi-2 pressure gauge; device passed all functional tests.
|
| |
|