| Model Number |
L-370 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No lot number was provided; therefore, device history record review could not be completed.A random sample of 32 units was taken from the manufacturing process, the units were visually inspected in order to verify that the components do not present any damage; no discrepancies were found during the inspection.The root cause was unable to be determined.As a preventive action, a notification of this complaint to the production personnel was conducted by document control technician.
|
| |
|
Event or Problem Description
|
|
It was reported the device routinely experienced the connective tubing (@ top of the l-370 sets) kinking off, when attached to either an iv admin.Set or blood set, in multiple areas.Multiple occasions, since began using hl-390 and l-370 over last few months.
|
| |
|
Search Alerts/Recalls
|