Ethnicity: requested, unknown, race: requested, unknown, explanted date: device was not explanted, occupation: lead technolegist.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.Review of the manufacturing record and the shipping inspection record of the involved product code/lot number combination confirmed that there were not any indications of anomaly in them.A search of the complaint file found no other similar report with the involved product code/lot number combination from other facilities.This product has a structure in which the stent self-dilates by rotating the thumbwheel (winding up the release wire connected to the sliding part) and pulling the sliding part toward the hand side.When the stent release is performed while the sliding part of delivery catheter is trapped, the movement of the sliding part on the hand side is hindered, and the stent is difficult to deploy.At that time, compressive force is applied to the proximal shaft located outside the patient's body, and waviness is occurred.The above may be prevented by properly holding the catheter.In the above state, when the operation of pushing and pulling the delivery catheter is performed, the force that has been compressed is released, the force that pushes out the inner shaft works, and the inner shaft (contrast marker) moves forward.The stent deployed from the sliding part (stent sheath) is pushed out by the proximal marker on the inner shaft to apply compressive force and is contracted.During stent deployment, when pulling force is applied to the delivery catheter to the hand side, an opening begins to occur in the deployed stent strut starting from the point where it is compressed at the stenosis part.Continued application of pulling force causes the stent strut to place in an elongated state.Based on our past knowledge, it was inferred that this event may have occurred by the following mechanism.However, since the actual product was not returned, it was not possible to clarify the cause of occurrence.The sliding part of actual product (stent mount section) was trapped by some hard object (e.G., calcified lesion).This prevented the sliding part from being pulled to the hand side, making deployment of the stent difficult.At that time, compressive force was applied to the proximal shaft located outside the patient's body, and waviness was occurred.Due to insufficient holding of the delivery catheter, operability declined with the continuation of the operation of rotating the thumbwheel, and the waviness of shaft progressed.The operation of pushing and pulling the delivery catheter was performed.Due to this, the force that had been compressed was released, the force that pushed out the inner shaft worked, and the inner shaft (contrast marker) moved forward.Then, the stent deployed from the sliding part (stent sheath) was pushed out by the proximal marker on the inner shaft to apply compressive force and was contracted.During stent deployment, when pulling force was applied to the delivery catheter to the hand side, an opening began to occur in the stent strut starting from the point where it was compressed at the stenosis part.As a result of continuing application of pulling force, the stent strut was placed in an elongated state.For future use, please keep in mind the following precaution and directions for use described in the instructions for use (ifu): · "[precautions] - pre-dilatation of the target vessel is recommended." · "[directions for use 3-6] - to maintain the position of the delivery catheter while rotating the thumbwheel, grip the delivery catheter by hand at the operator side (proximal) of the intermediate shaft and do not move the delivery catheter at the operator side of the intermediate shaft." · "[directions for use 3-6 precautions] - do not pull the delivery system tight during stent deployment.(the stent can become elongated during deployment." ashitaka factory manufacturing process has been making efforts in maintaining the quality of the product by performing following inspections.After the stent processing, 100% inspection is performed by an imaging device to assure that the stent strut is not deformed or fractured.After the stent processing, the width and thickness of the stent strut are inspected on 100% basis by an imaging device.After the stent is mounted on the delivery catheter, the length of stent is measured on 100% basis to assure it.After the stent is mounted on the delivery catheter, 100% visual inspection is performed to assure that there is no anomaly such as overlap, deformation or fracture of the stent strut.It is also confirmed that there is no kink or crush in the sliding part or the shaft.The load at the time of stent release is measured on 100% basis by the load measuring device to confirm that there is no anomaly in the release resistance.Terumo medical products (tmp) (importer) registration no.(b)(4) is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.(b)(4).Please see mdr 2243441-2022-00026 for the importer report.
|