• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported, during an unspecified procedure, an advance 35 lp low profile balloon catheter's balloon ruptured upon the first inflation attempt within a highly calcified lesion in the iliac artery.An unknown inflation device was used, as well as another manufacturer's 6 french catheter.The balloon catheter was removed by itself, and another device was used to successfully complete the procedure.The patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Customer name and address= line 2: (b)(6).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Event summary: as reported, during an unspecified procedure, an advance 35 lp low profile balloon catheter's balloon ruptured upon the first inflation attempt within a highly calcified lesion in the iliac artery.An unknown inflation device was used, as well as another manufacturer's 6 french catheter.The balloon catheter was removed by itself, and another device was used to successfully complete the procedure.The patient did not experience any adverse effects due to this occurrence.Investigation ¿ evaluation: reviews of the complaint history, device history record, instructions for use, drawing, specifications, and quality control procedures of the device were conducted during the investigation.The device was not returned to cook for investigation.The device history record records no relevant non-conformances.A database search for complaints on the reported lot found no additional complaints reported from the field.Cook concluded that no nonconforming product from this lot exists in house or in the field.The device is provided with instructions for use which warn, ¿do not exceed rated burst pressure.Rupture of balloon may occur.Adhere to balloon inflation pressure parameters in the compliance card insert.Over-inflation may cause rupture of the balloon, which resultant damage to the vessel wall.Use of a pressure gauge is recommended to monitor inflation pressures.¿ ¿do not use a power injector for balloon inflation or injection of contrast medium through catheter lumen marked ¿distal¿.Rupture may occur.¿ the instructions for use also states, ¿choose a balloon appropriate to lesion length and vessel diameter.¿ ¿upon removal from package, inspect the catheter to ensure no damage has occurred during shipping.¿ the instructions for use further notes - ¿note: if resistance is met while advancing the balloon dilatation catheter, determine the cause and proceed with caution.¿ ¿inflate balloon to desired pressure.Adhere to recommended balloon inflation pressures.¿ ¿if balloon pressure is lost and/or balloon rupture occurs, deflate balloon and remove balloon and sheath as a unit.¿ the device is provided with instructions for use which on how to supply, ¿store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ the information provided upon review of the dmr, dhr, and ifu suggests that there is evidence the device was manufactured to specification.Based on the available information, cook has concluded that the patient's calcified anatomy contributed to this incident.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per a review of risk documentation, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15638639
Report Number1820334-2022-01646
Device Sequence Number5881815
Product Code DQY
UDI-Device Identifier10827002523545
UDI-Public(01)10827002523545(17)250419(10)CINC001314
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K091527
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/19/2025
Device Model NumberN/A
Device Catalogue NumberPTA5-35-80-7-8.0
Device Lot NumberCINC001314
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/07/2022
Supplement Date Received by Manufacturer01/19/2023
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date01/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TERUMO 6 FRENCH CATHETER
Patient SexUnknown
-
-