• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number VP1332
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/16/2019
Event Type  malfunction  
Manufacturer Narrative
(udi) and g5 (510k) are unknown; no additional information was provided.This mdr was generated under protocol b10009704, as a result of warning letter cms# 617147.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Sample received consisted of one (1) product, in used conditions without its original packaging.The sample was visually inspected at a distance of 12 inches under normal lighting to received units, in order to detect any damage on the unit; it was found that the breathing bag is torn.The most probable root cause is that the production personnel did not detect and scrapped the damaged breathing bag.Action(s) taken to mitigate the reported issue(s): qalert was generated and production personnel was trained to ensure detection of any marks or damage on the breathing bags.
 
Event Description
It was reported that the breathing bag has sheared off at the circuit connector spot.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15638650
MDR Text Key306496491
Report Number3012307300-2022-25373
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/01/2021
Device Catalogue NumberVP1332
Device Lot Number3727581
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2021
Was the Report Sent to FDA? No
Date Manufacturer Received10/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age43 YR
Patient SexFemale
Patient Weight44 KG
-
-