Model Number BBP211
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 09/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information that during use of a pixie oxygenator, it was reported that there was a blood leak noticed from the top of the oxygenator during cardiopulmonary bypass for a patient with congenital heart disease.The device was replaced to complete the procedure.There was no additional patient impact associated with this event.Medtronic received additional information that there was no damage to the device, the packaging or other contents within the package (e.G.Devices in a tray or custom pack).The amount of patient blood loss and whether a transfusion was required were requested but this information could not be provided.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Correction section d expiration date: this field has been updated.Correction section h4 device manufacturing date: this field has been updated.Conclusion: the complaint was confirmed for the pixie oxygenator's leak from the top of the device.The issue was verified via the c ustomer-provided video, however, no device returned for analysis.The customer-provided video appears to show a blood leak near the antistatic shunt.The pixie oxygenator is 100% leak tested during production, and devices that exhibit leaks are discarded prior to distribution.Review of this unit¿s device history record found no abnormalities or non-conformance material report (ncmrs) initiated during manufacturing that would cause or contribute to the reported event.This device passed all leak testing and inspections during manufacturing.The root cause is undetermined.Trends for issues with this product are reviewed at quarterly quality meetings.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY PIXIE PEDIATRIC HOLLOW FIBER OXYGENATOR
Common Device Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15638713
Report Number2184009-2022-00277
Device Sequence Number20427242
Product Code DTZ
UDI-Device Identifier20613994617310
UDI-Public20613994617310
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K183511
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model NumberBBP211
Device Catalogue NumberBBP211
Device Lot Number223392453
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer11/10/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date12/06/2022
Date Device Manufactured12/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-