MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 15MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 15MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number FDS40015
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Codes Stroke/CVA (1770); Paresis (1998); Speech Disorder (4415)
Date of Event 09/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device is implanted inside the patient's vasculature.
 
Event or Problem Description
It was reported that the subject flow diverter stent was implanted successfully during the procedure with a minor ledge distally.Thirty minutes post procedure, the patient experienced weakness showing multiple stroke symptoms with weakness in the left side, facial droop and slurred speech prompting a return to angio suite for intervention.It was discovered that the proximal end of the subject flow diverter had collapsed.An additional surgery was performed, where the physician deployed an additional flow diverter stent to open the proximal end of the subject flow diverter stent.Patient is continuing to experience weakness and is still treated in the intensive care unit at the hospital but shows no stroke on imaging.Patient will continue on dapt and intergrelin for now.According to the physician the reported event was due to subject flow diverter stent failure on the proximal end.No further information is available.
 
Additional Manufacturer Narrative
D4 expiration date - added.H4 manufacturing date ¿ added.Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information indicated there was no anomalies to the fds prior to use, the dfu was followed and continuous flush was maintained throughout the procedure.Additional information states "30 minutes into recovery patient experience weakness prompting a return to angio suite.The patient was alert and responsive initially, but began showing multiple stroke symptoms with weakness in her left side, facial droop and slurred speech in pacu.During the assessment they called code stroke for her to return to the angio suite for intervention.Physician attempted to access the collapsed stent with a micro catheter to deploy an additional flow diverter stent to open the proximal end of the stent.Patient is still in the icu at the hospital with weakness but shows no stroke on imaging.Patient will continue on dapt and intergrelin".While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned to the reported issue ¿stent failed/unable to open¿.An assignable cause of anticipated procedural complication will be assigned to the reported ¿patient stroke¿, as a product related root cause does not apply and the issue is due to a known physiological effect of the procedure and/or patient condition noted with the directions for use, product labelling and/or risk documentation files.
 
Event or Problem Description
It was reported that the subject flow diverter stent was implanted successfully during the procedure with a minor ledge distally.Thirty minutes post procedure, the patient experienced weakness showing multiple stroke symptoms with weakness in the left side, facial droop and slurred speech prompting a return to angio suite for intervention.It was discovered that the proximal end of the subject flow diverter had collapsed.An additional surgery was performed, where the physician deployed an additional flow diverter stent to open the proximal end of the subject flow diverter stent.Patient is continuing to experience weakness and is still treated in the intensive care unit at the hospital but shows no stroke on imaging.Patient will continue on dapt and intergrelin for now.According to the physician the reported event was due to subject flow diverter stent failure on the proximal end.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURPASS EVOLVE 4.0MM X 15MM - PMAS
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15638746
Report Number3008881809-2022-00517
Device Sequence Number14923099
Product Code OUT
UDI-Device Identifier07613327375190
UDI-Public07613327375190
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P170024/S003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/16/2024
Device Model NumberFDS40015
Device Catalogue NumberFDS40015
Device Lot Number23110886
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer11/22/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR XT-27 CATHETER (STRYKER).; SCEPTER C 4X11 BALLOON (UNKNOWN).
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Other;
Patient Age71 YR
Patient SexFemale
-
-