• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 18D-SML; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 18D-SML; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE Back to Search Results
Model Number 55-0523
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Emotional Changes (1831); Foreign Body Reaction (1868); Pain (1994); Loss of Range of Motion (2032); Synovitis (2094); Anxiety (2328); Distress (2329); Discomfort (2330); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
Date of Event 02/03/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Litigation papers allege pain, injury, and elevated metal ions.Update rec'd (b)(6) 2015 partial medical records received.Patient reported to have low metal ion levels of co 1 and cr 1.1.Patient was admitted for revision to address discomfort, osteolysis and disability.Complete revision procedure reports were not received.The information received does not change the mdr decision.This complaint was updated on (b)(6) 2015.Updated rec'd (b)(6) 2016 supplemental received.Part and lot information was provided.There is no new information that would change the existing mdr decision.Complaint was updated (b)(6) 2016 update ad (b)(6) 2019: (b)(4) has been re-opened under (b)(4) due to medical records and sticker sheets received.After review of medical records, the patient was revised due to painful tha of left hip, mechanical complication and metal debris induced synovitis, hip effusion and osteolysis of the femur.Operative note reported a large amount of brownish colored fluid, there is a mild amount of corrosion under the taper, fibrinous brown material in the joint with debris and brownish stained synovium, lysis in the greater trochanter and missing bone.Operative finding reported osteolysis in the femur and fluid pocket in the joint.Added expiration date , patient harm, pe code of cup and head, medical history and revision hospital.Added new ip stem due to a corrosion and osteolysis.Doi: (b)(6) 2006: dor: (b)(6) 2015; (left hip).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15638791
Report Number1818910-2022-20990
Device Sequence Number20304067
Product Code LPH
UDI-Device Identifier10603295175285
UDI-Public10603295175285
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number55-0523
Device Catalogue Number550523
Device Lot Number1955380
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer12/26/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date12/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/18/2005
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADAPTER SLEEVE 11/13 +6; ASR ACETABULAR CUPS 56; ASR UNI FEMORAL IMPL SIZE 49; SROM STM STD 36+12L 13X18
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexMale
Patient Weight71 KG
-
-