CLEARCORRECT CLEARCORRECT; CLEAR PLASTIC ORTHODONTIC ALIGNER Back to Search Results
Model Number Aligner (Multi-layer)
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Difficulty Chewing (4522)
Date of Event 08/05/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Clearcorrect findings: a review of the case ledger shows numerous delays/extended timeframes during treatment.The initial impressions were submitted on march 8, 2019 and a setup was sent on (b)(6) 2019.The version 2 setup was approved on (b)(6) 2019.The 12-step phase 1 shipped on june 15, 2019 with a two-week wear schedule, the patient should have completed the treatment by late (b)(6) 2019.A revision was submitted on february 21.2020 2020; the version 3 treatment was approved on (b)(6) 2020, and the 5-step phase 2 shipped in march 15, 2020.Another revision was submitted on june 25, 2020, but scans were not received until august 19, 2020 and the version 4 treatment was not approved until (b)(6) 2020.The 5-step phase 3 was shipped on september 15, 2020.The treatment should have completed by late (b)(6) 2020, but the last revision was not submitted until march 31, 2021, and the records were received on april 7, 2021.The version 5 setup was approved on (b)(6) 2021, and the 5-step phase 4 was shipped on may 21, 2021.The treatment should have completed by august of 2021, but the retainer based on existing records was not ordered until (b)(6) 2021.A comparison of the last step from the phase 3 (step 23) to the scans received on april 7, 2021 shows signs of tracking issues and incomplete movements.Unable to determine any treatment planning contributing factors that would lead to tracking issues.All treatment planning protocols and prescriptions were followed.Clinical evaluation: the etiologies of temporomandibular joint dysfunction (tmd) encompass a broad range of factors which include occlusal abnormalities, orthodontic treatment, bruxism and orthopedic instability, macrotrauma and microtrauma, joint laxity and exogenous estrogen.Psychological factors such as stress, mental tension, anxiety or depression can cause tmd.Within the structures of the temporomandibular joint are hard and soft tissues which may be affected.It is noted that the patient had seen a specialist who diagnosed myofascial pain and internal derangement and was managed with a splint.Neither of these diagnosis are reported to be associated with aligner therapy.As such, it is most likely that the patient's symptoms are coincident with aligner therapy as opposed to the result of it.Regulatory evaluation: this case is related to treatment provided by private-label customer candid; as legal manufacturer, (b)(4) is responsible for adverse event reporting.We are assessing this event as we are the physical manufacturer of the devices used in this case, so this is relevant to our post-market surveillance activities.Based upon our investigation, there appears to have been patient compliance issues throughout this case.A number of revisions were needed to complete the treatment.Ultimately a retainer was ordered based upon the final treatment step; ordering a retainer is an indication that the patient and clinician were satisfied with the outcome of aligner therapy.The retainer was ordered (b)(6) 2021.The information provided to clearcorrect by candid indicates that the patient first complained about their "bite being off" in (b)(6) 2021 (this coincides with receipt of the phase 2 devices).The patient reported to candid that they had seen a specialist on (b)(6)2021 who had prescribed a jaw splint.Candid did not report this event to clearcorrect until (b)(6) 2022.It is unclear whether the patient may have had issues with tmj prior to the start of treatment.Because this event occurred during the time of aligner treatment, it is possible that the treatment was a contributing factor.The patient did seek medical treatment to address the "bite being off." the patient was prescribed a jaw splint, which is a medical intervention in this case.Based upon these details, this event does meet the criteria for adverse event reporting.
 
Event or Problem Description
Per the candid co, the patient reports tmj issues, and that they've seen a specialist that prescribed a jaw splint.The patient first reported their bite being off in (b)(6) 2021, during treatment.However, bite issues are a known and disclosed side effect of all orthodontic treatment.On (b)(6) 2021, the patient reported that they'd seen the jaw specialist and required a jaw splint.A product return is not possible, because the patient alleges that treatment, overall, has caused this issue, and not a specific step, arch, or device.There are no reports of the device being altered and no photos of the devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARCORRECT
Common Device Name
CLEAR PLASTIC ORTHODONTIC ALIGNER
Manufacturer (Section D)
CLEARCORRECT
21 cypress blvd
round rock TX 78665
Manufacturer (Section G)
CLEARCORRECT
21 cypress blvd
round rock TX 78665
Manufacturer Contact
raquel brown
21 cypress blvd
round rock, TX 78665
8883313233
MDR Report Key15638794
Report Number3007130440-2022-00010
Device Sequence Number5867911
Product Code NXC
UDI-Device IdentifierD865REF00031
UDI-Public+D865REF00031/$$522074269534/16D20200915
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K113618
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Dentist
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date03/15/2022
Device Model NumberAligner (Multi-layer)
Device Catalogue NumberREF-03
Device Lot Number269534
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2022
Initial Report FDA Received Date10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age30 YR
Patient SexMale
-
-