| Model Number |
IBI-89003 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the premature ventricular contractions procedure, the cool point pump would unintentionally power off and needed to be restarted 6-8 times during the procedure resulting in a delay.The catheter was exchanged with a safire 4mm dry tip ablation catheter until another cool point pump could be used.The dry tip rf lesions were reinforced with the tacticath se ablation catheter with irrigation using a spare cool point pump.The procedure was successfully completed with no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
One cool point irrigation pump was received for complaint evaluation.The pump functioned normally during the functional testing and the servicing.No issue was observed.The pump was left running for half an hour, the power did not off.The device met specifications prior to release from abbott manufacturing facilities as supported by the device history record.
|
| |
|