It was reported that during the surgery, could not fire without motor sound on the 1st firing.The surgeon removed the battery and rotated for 180 degree and reinserted the battery.The device still could not fire and without motor sound.Changed another one to continue the surgery, after fired, noted the staples malformed.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
|
|
(b)(4).Batch #: 286a31.Additional information: this is an analysis of a photo submitted to for evaluation.During the visual analysis, the following was observed: the photo shows a staple line on the tissue and some staples leg not properly formed were noted on the tissue.No conclusion or root cause could be determined.Please refer to the device analysis for full analysis details and conclusion.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pce60a device was returned with no apparent damage and with no reload present.The manual override door was noted to be out of position; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.The bailout system was reset and then, it was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: tissue thickness should be carefully evaluated before firing any stapler.Holding the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation.When positioning the stapler on the application site, ensure that no obstructions such as clips, stents, guide wires, etc.Are within the instrument jaws.The event described could not be confirmed as the device performed without any difficulties noted.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
|