• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIRONA DENTAL SYSTEMS GMBH MIDWEST PHOENIX; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIRONA DENTAL SYSTEMS GMBH MIDWEST PHOENIX; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 791400
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
3 of 3 devices were returned for evaluation.Evaluation of 3 devices found normal chuck wear and the turbines needed to be replaced.The devices were repaired and returned to the customers.
 
Event Description
This report summarizes 3 malfunction events where a midwest phoenix handpiece would not hold burs.No injury resulted in any of the events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST PHOENIX
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494593
MDR Report Key15639403
MDR Text Key306593342
Report Number9614977-2022-00073
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K131319
Number of Events Reported3
Summary Report (Y/N)Y
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number791400
Patient Sequence Number1
-
-