| Model Number |
8884721252E |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
Unspecified Mental, Emotional or Behavioural Problem (4430)
|
| Date of Event |
10/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported it was difficult, even impossible to remove the guide, even after lubricating the tube and insertion.The product was checked before use and it was adequate.They tried another tube and had the same problem.There was trauma for the beneficiary.Per additional information provided on october 19, 2022 there was no physical trauma but rather psychological trauma related to the repeated attempts to withdraw the guide wire from the tube as well as the back and forth movements of the tube in the nasal wall, the throat, and in the esophagus during attempts to withdraw the guide with the associated risks.The tube was finally removed and another tube was installed with the same difficulties.To their knowledge there has been no medical treatment/intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured on december 20, 2021.The investigation was carried out with the multifunctional team.A gemba walk was performed, and all processes and controls were found to be followed correctly, including subassemblies, finished product assembly, packaging, and product inspections.No abnormal conditions were found that could trigger the reported condition.One sample was received at the manufacturing site and evaluated; the hydromer was activated according to the instructions for use for proper procedure for insertion of the feed tube and extraction of the stylet, ¿using a syringe, inject approximately 10 cc of water into the tube to activate the hydromer coating¿.The reported condition could not be confirmed, the stylet was extracted from the tube and the product was confirmed to meet specifications.A corrective action is not applicable at this time.We will continue to monitor customer complaints and feedback notifications for adverse trends that require immediate attention.This complaint will be used for tracking and trending purposes.
|
| |
|
Search Alerts/Recalls
|