• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW 7 XB BISECTOR (US); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW 7 XB BISECTOR (US); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEW 7 XB BISECTOR (US)
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Trackwise id (b)(6).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event or Problem Description
The hospital reported that during an endoscopic vein harvesting procedure the vasoview 7 xb bisector for the vein and the knife would not fully extend making it impossible to get around branches or use it without damaging the vein.They inspected it and cleaned it, but it appeared the knife was bent and off its track.They opened a new kit for the bisector and that one worked correctly.No harm or risk of harm to the patient.No delay.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Rackwise #(b)(4).The device was returned to the factory for evaluation on 10/05/2022.An investigation was conducted on 10/17/2022.A visual inspection was conducted.Signs of clinical use and evidence of blood was observed on the harvesting device handle.The bipolar electrodes were observed to be intact, with no visual defects observed.The cutter was to be intact, with no visual defects observed.A mechanical evaluation was conducted.The blue toggle was manipulated to extend and retract the cutter blade.There were no visual or physical difficulties observed.Based on the returned condition of the device as well as the evaluation results, the reported failure "material twisted/ bent" was not confirmed.The lot history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW 7 XB BISECTOR (US)
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key15639984
Report Number2242352-2022-00866
Device Sequence Number6376895
Product Code GEI
UDI-Device Identifier00607567700369
UDI-Public00607567700369
Combination Product (Y/N)N
Initial Reporter StateNY
PMA/510(K) Number
K041981
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/11/2024
Device Model NumberVASOVIEW 7 XB BISECTOR (US)
Device Catalogue NumberVH-3200
Device Lot Number3000255555
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer10/21/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-