STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICM5_12.6
Medical Device Problem Codes Misfocusing (1401); Off-Label Use (1494)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 09/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter indicated that a 12.6mm vicm5_12.6, -8.0 diopter implantable collamer lens was implanted into the patient's left eye (os) on (b)(6) 2022.It was reported that limbal relaxing incisions (lris) were made to reduce astigmatism but it did not reduce the expected amount.On (b)(6) 2022 the lens was exchanged for a toric lens to address the astigmatism (high cylinder power) and the problem resolved.In the reporters opinion the cause of this event was user error, product did not fail to perform as intended.
 
Additional Manufacturer Narrative
Additional information: h6: device code 1494- off label use- patient acd<3.00mm.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
8002927902
MDR Report Key15640127
Report Number2023826-2022-03703
Device Sequence Number6444540
Product Code MTA
UDI-Device Identifier00841542119146
UDI-Public00841542119146
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model NumberVICM5_12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer11/11/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date12/23/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45 W/FTP, LOT# UNK; INJECTOR MODEL: MSI-PF, LOT# UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexFemale
-
-