SYNTHES GMBH 1.1MM DRILL BIT/K-WIRE ATTCHMT FOR THREADED HOLE/70MM; BIT, DRILL Back to Search Results
Catalog Number 03.130.200
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2022, the patient underwent the open reduction internal fixation with the drill bits and the screws for the comminuted fracture of the proximal phalanx and the index proximal phalanx.When the surgeon drilled to the plate without a drill sleeve, the tip of the drill bit broke and remained in the patient¿s body.It took approximately 30 minutes to remove the broken tip.The surgeon drilled with a new drill bit and a drill sleeve, and the tip of the drill bit broke and remained in the patient¿s body again.It took more than 30 minutes.Both drills were broken at the border of the flute.It was confirmed by x-rays that there were no broken pieces in the patient¿s body.In addition, when the surgeon was inserting the screws, the screws stopped in the middle of the road due to wrong direction.Although the surgeon removed the screws and re-inserted them several times, the screw heads broke and could not be removed.Therefore, the surgeon removed the screws with a plier.It was confirmed by x-rays that there were no broken screws in the patient¿s body.The surgery was completed successfully with two hours delay.The operation was scheduled to take two hours; however, it took four hours.According to the surgeon, the screws direction might have been at an angle to the drilled hole.The surgeon also commented that he tried to insert the screws with a driver forcefully although the patient had good bone quality which might cause the event.This report involves one (1) 1.1mm drill bit/k-wire attchmt for threaded hole/70mm.This is report 2 of 6 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Therapy date: (b)(6) 2022.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Part: 03.130.200, lot: f-23440, manufacturing site: selzach, supplier: (b)(4), release to warehouse date: 18.Jan.2018.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the tip of the drill bit ø1.1 l70/39 f/core hole, p/n: 03.130.200, broke off.The broken fragment was returned for examination.No other problems identified.A dimensional inspection was unable to be performed due to not being applicable to the complaint condition.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the drill bit ø1.1 l70/39 f/core hole, p/n: 03.130.200 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.1MM DRILL BIT/K-WIRE ATTCHMT FOR THREADED HOLE/70MM
Common Device Name
BIT, DRILL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15640154
Report Number8030965-2022-08518
Device Sequence Number15046216
Product Code HTW
UDI-Device Identifier07611819690585
UDI-Public07611819690585
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.130.200
Device Lot NumberF-23440
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/29/2022
Supplement Date Received by Manufacturer11/14/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/18/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORTSCR Ø1.5 SELF-TAP L12 TAN; DRILL BIT Ø1.1 L70/39 F/CORE HOLE; VA LOCKSCR Ø1.5 SELF-TAP L11 TAN; VA LOCKSCR Ø1.5 SELF-TAP L12 TAN; VA LOCKSCR Ø1.5 SELF-TAP L14 TAN
Patient SexUnknown
-
-