MAUDE Adverse Event Report: MEGADYNE MEDICAL PRODUCTS, INC. 10FT SMOKE PENCIL UCONN; RETURN ELECTRODE CABLE, DUAL PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEGADYNE MEDICAL PRODUCTS, INC. 10FT SMOKE PENCIL UCONN; RETURN ELECTRODE CABLE, DUAL PLATE Back to Search Results
Model Number 251010J
Medical Device Problem Codes Flaked (1246); Difficult or Delayed Activation (2577)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 10/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot or batch number was provided therefore a device history could not be done.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: are there any photos of the burn that you could share with us in regards to the burn? if yes, please send to productcomplaint1@its.Jnj.Com.Are there any photos of the device that you could share with us in regards to the burn? if yes, please send to productcomplaint1@its.Jnj.Com.What is the severity of the burn? (please see degrees of burns below and choose one) first degree burns are minor burns on the first layer of skin.The skin looks dry, redness, and may be swelling; no penetration or blisters.Second degree burn looks wet or moist.The burn site appears red, blistered, and may be swollen and painful.Third degree burn the burn site looks deep, whitening or blackened and charred what medical intervention was used to treat the burn? (such as salve or stitches).Besides the burn, did the patient experience any adverse consequence due to the issue? are there any anticipated long-term effects from the burn or injury? what is the current status of the patient? does the surgeon believe there is there an alleged deficiency to the device that led to patient burn and if so why? what generator was being used? what power levels was generator set to? what was the return electrode that was being used? what does the surgeon believe was the cause of the surgical drape to catch on fire? what mode was the generator set to? can you describe how the endofactor was cleaned throughout the procedure? where was the location of the burn? what was the indication for surgery? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during the excision of temporal mass procedure, the surgical drape caught fire.It was unknown how the procedure was completed.The patient was moved to a burned hospital and is now in icu.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 11/18/2022 investigation summary an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.Additional information was requested, and the following was obtained: what generator was being used? valleylab/medtronic what was the return electrode that was being used? medtronic all other information is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10FT SMOKE PENCIL UCONN
Common Device Name
RETURN ELECTRODE CABLE, DUAL PLATE
Manufacturer (Section D)
MEGADYNE MEDICAL PRODUCTS, INC.
4545 creek road
cincinnati OH 45242
Manufacturer (Section G)
MEGADYNE MEDICAL PRODUCTS, INC.
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key15640285
Report Number1721194-2022-00094
Device Sequence Number6376966
Product Code GEI
UDI-Device Identifier10614559105573
UDI-Public10614559105573
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K982130
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number251010J
Device Catalogue Number251010J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer11/18/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-