• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNITED STATES ENDOSCOPY GROUP, INC. CARR-LOCKE INJECTION NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNITED STATES ENDOSCOPY GROUP, INC. CARR-LOCKE INJECTION NEEDLE Back to Search Results
Model Number 00711811
Medical Device Problem Codes Loss of or Failure to Bond (1068); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported via medwatch report (b)(4) that during a patient procedure, the handle to their carr-locke injection needle device came apart and was not operating properly.No report of injury.
 
Additional Manufacturer Narrative
Steris requested the carr-locke injection needle device back for evaluation.The information for use packet (ifu 730700) provides the following information when using the articulator¿ - injection needle and the carr-locke injection needles, see below."carefully remove the tip protector from the distal tip of the needle.Actuate the proximal luer back and forth, visually observing the needle being adequately advanced from the device's distal end.The spring-loaded handle assists in needle retraction; however, always visually check the retraction as well.Note: do not stretch the device because stretching may cause the needle projection length to be altered.Do not attempt to actuate the injection needle with the catheter in a straightened position.The injection needle has been engineered and manufactured with a specific "preload" that is necessary to assure the full projection and full retraction of the needle from the distal hub when the device is in very tortuous configurations." the device history record was reviewed and confirmed the lot subject of the event was manufactured to specification.Investigation of this event is currently in process.A follow-up report will be submitted when additional information becomes available.
 
Additional Manufacturer Narrative
The device subject of the reported event was returned for evaluation.The evaluation determined the handle components had separated due to insufficient adhesive at the connection between the components.A review of complaints determines this to be an isolated event.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARR-LOCKE INJECTION NEEDLE
Common Device Name
INJECTION NEEDLE
Manufacturer (Section D)
UNITED STATES ENDOSCOPY GROUP, INC.
5976 heisley road
mentor OH 44060
Manufacturer (Section G)
UNITED STATES ENDOSCOPY GROUP, INC.
5976 heisley road
mentor OH 44060
Manufacturer Contact
coletta cohara
5976 heisley road
mentor, OH 44060
4403586251
MDR Report Key15640442
Report Number1528319-2022-00053
Device Sequence Number12391799
Product Code FCG
UDI-Device Identifier00724995180881
UDI-Public00724995180881
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00711811
Device Catalogue Number00711811
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-