• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ON-X LIFE TECHNOLOGIES, INC. ON-X ASCENDING AORTIC PROSTHESIS WITH THE VASCUTEK GELWEAVE VALSALVA GRAFT; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ON-X LIFE TECHNOLOGIES, INC. ON-X ASCENDING AORTIC PROSTHESIS WITH THE VASCUTEK GELWEAVE VALSALVA GRAFT; HEART-VALVE, MECHANICAL Back to Search Results
Model Number ONXAAP-25
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pseudoaneurysm (2605)
Date of Event 04/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This investigation is currently ongoing.Any additional information will be provided in the follow-up report.This report is being submitted as required by federal regulations and does not constitute an admission that the device caused or contributed to the reported event.Furthermore, this report reflects the event as alleged by the complainant and does not imply that the information reported to artivion- formerly cryolife/jotec is accurate or has been confirmed by artivion.
 
Event or Problem Description
According to the initial report received, study patient experienced a pseudoaneurysm and underwent reoperation on (b)(6), 2022, 16-days following implant with the device.
 
Additional Manufacturer Narrative
The manufacturing records were reviewed and it was confirmed that all records were controlled, available for review, and met all specifications per the device master record.All lots passed functional testing and met release specifications.During the investigation no non-conformances or deviations were found to be related to the complaint.Device was implanted on (b)(6) 2022 to treat a type a thoracic aortic aneurysm without rupture in a subject of unknown age and sex.The subject was enrolled in study.On (b)(6) 2022 (16 days post implant) the subject complained of chest pain and a ct angiogram was performed.It showed ¿pseudoaneurysm in the cranial/right portion which corresponds to the suture of the common trunk button and significant uptake of the contrast which may correspond to inflammation/infection.¿ a second surgical revision was performed and fresh periaortic hematoma and bleeding on the posterior side the prosthesis with observation of dissection of the button of the left coronary ostium.The treatment of a suture and a pericardium patch were applied; cultures were taken with negative results.At the time of discharge an echocardiogram and chest xray were performed.The echo noted ¿mechanical bentall with prosthesis with no signs of dysfunction.Remains of the periaortic hematoma still present in the annulus and around the tube in the proximal ascending aorta.Minimal pericardial effusion in the anterior sac, insignificant, without sign of hemodynamic compromise.The chest xray noted ¿no infiltrates or consolidations of the pulmonary parenchyma and no signs of pleural effusion.¿ the event was ¿resolved without sequelae¿ and the subject was later discharged home.There is no evidence that the valve contributed to the need for the reoperation.Nevertheless, the instructions for use [ifu] note that reoperation is a risk associated with the implantation of an aap device.The ifu also lists false aneurysm as a potential adverse event, which, in this case, precipitated the reoperation.The risk management and usability engineering file was reviewed and thoroughly identifies the process and product hazards for approved indications.Each individual hazard is mitigated and reduced as far as possible by design and process.Post production residual risk is communicated in the product¿s labeling and ifu (instructions for use).No new risks were identified during the course of the review.This report is being submitted as required by federal regulations and does not constitute an admission that the device caused or contributed to the reported event.Furthermore, this report reflects the event as alleged by the complainant and does not imply that the information reported to artivion- formerly cryolife/jotec is accurate or has been confirmed by artivion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-X ASCENDING AORTIC PROSTHESIS WITH THE VASCUTEK GELWEAVE VALSALVA GRAFT
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ON-X LIFE TECHNOLOGIES, INC.
1300 e. anderson ln., bldg. b
austin TX 78752
Manufacturer (Section G)
ON-X LIFE TECHNOLOGIES, INC.
1300 e. anderson ln., bldg. b
austin TX 78752
Manufacturer Contact
rochelle maney
1655 roberts blvd. nw
kennesaw, GA 30144
7704193355
MDR Report Key15640604
Report Number1649833-2022-00051
Device Sequence Number15061967
Product Code LWQ
UDI-Device Identifier00851788001532
UDI-Public851788001532
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P000037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberONXAAP-25
Device Catalogue NumberONXAAP-25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2022
Supplement Date Received by Manufacturer09/29/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date12/16/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-