Updated fields: b4,d1,d9,e1(site country),g3,g6,h2,h3,h4,h6(type of investigation, investigation findings, component codes, investigation conclusions), h10 a getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.To fix the issue, the fse replaced the safety disc locking lap , and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released for cleared for clinical service.
|