| Model Number |
N/A |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the unit does not power on.The event timing was outside of surgery.There was no harm or delay reported.No adverse events were reported as a result of this malfunction upon investigation, there were signs of charring on the power supply board.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the cart would not power on; the power supply shorted.The power supply was replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|