• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSTRUMENTATION LABORATORY CO. GEM PREMIER 4000; CLINICAL CHEMISTRY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSTRUMENTATION LABORATORY CO. GEM PREMIER 4000; CLINICAL CHEMISTRY INSTRUMENT Back to Search Results
Model Number 2261
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
Customer submitted a medwatch user facility report (b)(4) that indicated that a gem premier 4000 experienced four consecutive cartridge failures.After the reported cartridge failures, werfen service replaced the heater assembly which returned the instrument to operational condition.Date of event indicated (b)(6) 2022.
 
Additional Manufacturer Narrative
The user facility report indicated that the service was completed following the failure of four cartridges.Service took place on (b)(6) 2022, between the third and fourth cartridge failures.A review of complaints identified four cartridge failure complaints.Two cartridge failures occurred on (b)(6) 2022, a third on (b)(6) 2022 and a fourth on (b)(6) 2022.Service was completed on (b)(6) 2022, during which the instrument's heater assembly was replaced and all system checks were completed; the instrument was fully operational and ready for normal use.Investigation of the complaints revealed that reported issues were related to individual cartridge issues and not the instrument.The first cartridge failure was due to error 2016 (ground voltage error).The subsequent cartridge failures were due to process control solution not detected.There was no known patient impact.Unable to determine the cause of the cartridge failures.The instrument operated as intended.There is no indication of a product quality issue with respect to manufacture, design, or labeling.As a precaution, the heater assembly was replaced on the instrument.However, investigation did not determine that the cause of the issue was instrument related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEM PREMIER 4000
Common Device Name
CLINICAL CHEMISTRY INSTRUMENT
Manufacturer (Section D)
INSTRUMENTATION LABORATORY CO.
180 hartwell road
bedford MA 01730
Manufacturer (Section G)
INSTRUMENTATION LABORATORY CO.
180 hartwell road
bedford MA 01730
Manufacturer Contact
reba daoust
180 hartwell road
bedford, MA 01730
MDR Report Key15642141
Report Number1217183-2022-00007
Device Sequence Number12356643
Product Code CHL
UDI-Device Identifier08426950452591
UDI-Public08426950452591
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K133407
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number2261
Device Catalogue Number00025000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/21/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-