• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS MEXICO ONE-STEP¿ CENTESIS CATHETER; CATHETER, IRRIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS MEXICO ONE-STEP¿ CENTESIS CATHETER; CATHETER, IRRIGATION Back to Search Results
Model Number 00884450092712
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that during use, the device was found to be leaking.No additional patient consequence to report.
 
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and 10 similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONE-STEP¿ CENTESIS CATHETER
Common Device Name
CATHETER, IRRIGATION
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS MEXICO
8830 siempre viva rd #100
san diego CA 92154
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS MEXICO
8830 siempre viva rd #100
san diego CA 92154
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
MDR Report Key15642374
Report Number3011642792-2022-00101
Device Sequence Number12373168
Product Code GBX
UDI-Device Identifier00884450092712
UDI-Public00884450092712
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/26/2024
Device Model Number00884450092712
Device Catalogue NumberOSC-5F-7LT/B
Device Lot NumberI2407523
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer12/04/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-