• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL PNEUPAC VENTILATORS VENTIPAC; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL PNEUPAC VENTILATORS VENTIPAC; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Model Number 130003
Medical Device Problem Codes Device Alarm System (1012); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the battery alarm is not working properly.No patient involved.
 
Additional Manufacturer Narrative
A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Operator of device is unknown.
 
Additional Manufacturer Narrative
Other, other text: h6: event problem and evaluation codes: updated.H3: device evaluated by manufacturer: updated h10: device evaluation: the device was returned for investigation.Visual inspection and functional test were performed.The customer reported problem was not related to any previous repair.Visual inspection found the device with no physical damage.The tamper seals and labels were intact.There was no evidence of the error recorded in the event history log.The customer reported problem was verified/duplicated during functional testing.Performed functional test and verified the light-emitting diode (led) display was not working properly (led failure) with constant low battery alarm.The most probable cause was due to a faulty main board.The root cause was undetermined/unknown.For corrective action, the faulty main board was replaced.The previous service history has been reviewed and deemed unrelated to the current customer reported problem.This remediation mdr was generated under protocol b10010579, as a result of warning letter cms# 617147.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC VENTILATORS VENTIPAC
Common Device Name
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
receiving
minneapolis, MN 55442
MDR Report Key15642414
Report Number3012307300-2022-25525
Device Sequence Number6435949
Product Code BTL
UDI-Device Identifier10610586038259
UDI-Public10610586038259
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K030803
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number130003
Device Catalogue Number130003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2022
Supplement Date Received by Manufacturer05/22/2023
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date06/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2006
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-