• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC ACROBAT-I POSITIONER; STABILIZER, HEART

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC ACROBAT-I POSITIONER; STABILIZER, HEART Back to Search Results
Model Number ACROBAT-I POSITIONER
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that during a coronary artery bypass procedure, acrobat-i positioner knob didn't engage and spun multiple times without locking in.No patient affects and they used the device to finish the case.
 
Additional Manufacturer Narrative
Trackwise id (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Additional Manufacturer Narrative
Trackwise#: (b)(4).Corrected section: d4 corected from "(b)(4)" to "(b)(4)".The lot # 3000199248 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.H3 other text: device not returned.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT-I POSITIONER
Common Device Name
STABILIZER, HEART
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key15642584
Report Number2242352-2022-00868
Device Sequence Number6435963
Product Code MWS
UDI-Device Identifier00607567500006
UDI-Public00607567500006
Combination Product (Y/N)N
Initial Reporter StateGA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/18/2023
Device Model NumberACROBAT-I POSITIONER
Device Catalogue NumberC-XP-5000Z
Device Lot Number3000199248
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2022
Supplement Date Received by Manufacturer01/10/2023
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
Patient Weight89 KG
-
-