Model Number 647205
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using the bd facs¿ sample prep assistant iii that the washing column of the probe is overflowing.The following information was provided by the initial reporter: the customer reports that the washing column of the probe is overflowing.The customer cleaned the filter (which was clean) but the problem persists.
 
Additional Manufacturer Narrative
Common device name: automated pipetting, diluting and specimen processing workstations for flow cytometric analysis.(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
This pr is made to capture the correct instal date.The correct install date is (b)(6) 2017.
 
Event or Problem Description
It was reported that while using the bd facs¿ sample prep assistant iii that the washing column of the probe is overflowing.The following information was provided by the initial reporter: the customer reports that the washing column of the probe is overflowing.The customer cleaned the filter (which was clean) but the problem persists.
 
Event or Problem Description
It was reported that while using the bd facs¿ sample prep assistant iii that the washing column of the probe is overflowing.The following information was provided by the initial reporter: the customer reports that the washing column of the probe is overflowing.The customer cleaned the filter (which was clean) but the problem persists.
 
Additional Manufacturer Narrative
H.6 investigation summary (b)(4) pn: 647205 spaiii sn: (b)(6) awareness: 10/13/2022 opened: 10/13/2022 investigation summary: ¿ scope of issue: the scope of issue is limited to part: 647205 spaiii and serial number: (b)(6) ¿ problem statement: customer reported: overflow washing column ¿ manufacturing defect trend: there are 0 qns related to the reported issue.Date range (date of incident to 12 months back) from 13oct2021 to date 13oct2022 (rolling 12 months) ¿ complaint trend: there are 10 complaints related to the reported complaint.Date range (date of incident to 12 months back) from 13oct2021 to date 13oct2022 (rolling 12 months) o complaint data attached ¿ investigation result / analysis: per fse¿s report: general check of the instrument carried out.O replaced the waste miniwash pump assembly ((b)(6)).O performed alignment and tests with complete rack - passed o checked for correct operation o no further issues o sample preparation was not affected.O there was no delay in patient treatment due to any unexpected results.¿ service max review: review of related work order#(b)(4) install date: (b)(6) 2008 defective part number: 334297 ¿ miniwash assembly work order notes: o subject / reported: overflow washing column o problem description: wash tower is leaking o cause: wash pump assembly o work performed: replaced miniwash assembly o solution: replaced miniwash assembly o parts replaced: 334297 ¿ miniwash assembly ¿ returned sample analysis: the complaint sample was not requested to be returned and no photo(s) / picture(s) were provided ¿ manufacturing device history record (dhr) review: review of the dhr for serial number: (b)(6) and pn647205 was reviewed.The instrument met all the manufacturing specifications prior to release.¿ risk analysis: o risk management file part #100245ra, revision 03 was reviewed.O hazard(s) identified? yes no hazard id: 3.1.29 hazard: environment biohazard cause: wash station backs up due to waste pump failure harmful effects: exposure to biohazard residual severity: 5 residual probability: 1 residual risk index: 5 ¿ root cause: based on the investigation result and fse¿s report the root cause was a defective waste pump ¿ conclusion: based on the investigation results and the fse¿s report the complaint was confirmed for overflowing wash column (station).No one was harmed or injured, and no medical diagnosis was performed due to any incorrect results.There was no impact to the patient health or safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD FACS¿ SAMPLE PREP ASSISTANT III
Common Device Name
SEE H.10
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15642814
Report Number2916837-2022-00317
Device Sequence Number2748652
Product Code PER
UDI-Device Identifier00382906472052
UDI-Public00382906472052
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K131301
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number647205
Device Catalogue Number647205
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2022
Supplement Date Received by Manufacturer12/19/2022
01/31/2023
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date12/21/2022
02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/14/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1