• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON AND CO. BD NANO¿ ULTRA-FINE¿ PEN NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON AND CO. BD NANO¿ ULTRA-FINE¿ PEN NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 8975
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/26/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd nano¿ ultra-fine¿ pen needle broke off in the injection site.The following information was provided by the initial reporter: consumer reported that the needle broke off in the injection site.
 
Additional Manufacturer Narrative
H6: investigation summary: no samples were returned therefore the complaint could not be confirmed and the root cause is undetermined.A lot history review was carried out and no related non conformances were raised in association with this packaged lot concluding all inspections were performed as per the applicable operations and met qc specifications.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd nano¿ ultra-fine¿ pen needle broke off in the injection site.The following information was provided by the initial reporter: consumer reported that the needle broke off in the injection site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NANO¿ ULTRA-FINE¿ PEN NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON AND CO.
pottery road
dun laoghaire co
Manufacturer (Section G)
BECTON DICKINSON AND CO.
pottery road
dun laoghaire co
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15643147
Report Number9616656-2022-01118
Device Sequence Number6194062
Product Code FMI
UDI-Device Identifier10885403897511
UDI-Public(01)10885403897511
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K213478
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number8975
Device Catalogue Number320883
Device Lot Number1351576
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer11/09/2022
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-