• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381437
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd insyte¿ autoguard¿ shielded iv catheter needle would not retract.This complaint was created to capture the 2nd of 2 related incidents.The following information was provided by the initial reporter: "we have had 2 instances where the above needle did not retract.".
 
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd insyte¿ autoguard¿ shielded iv catheter needle would not retract.This complaint was created to capture the 2nd of 2 related incidents.The following information was provided by the initial reporter: "we have had 2 instances where the above needle did not retract.".
 
Additional Manufacturer Narrative
H.6.Investigation summary: since no samples displaying the condition reported were available for examination, we were unable to fully investigate this incident.Examination of the actual product involved may provide clarification as to the cause for the reported failure.A device history record review showed no non-conformances associated with this issue during the production of this batch.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 07-dec-2022.H6: investigation summary our quality engineer inspected the sample submitted for evaluation.Bd received one used 20g x 1.88in.Insyte autoguard unit from lot number 1300906.A gross visual inspection of the returned unit found that the needle was in the un-retracted position and the button had not been depressed.The catheter assembly was not returned.Further microscopic inspection found adhesive between the needle hub and grip which prevented needle retraction.The reported issue was confirmed.This was physical/mechanical evidence to confirm and support a manufacturing process related issue for the reported defect relating to the adhesive dispensing station.This may occur at the adhesive dispense process due to station misalignment, part misalignment or adhesive buildup on the nozzle.Per our quality control plan, inspections for needle retraction and excessive adhesive are performed periodically during the manufacturing process to mitigate the occurrence of this defect.Preventative maintenance (pm¿s) is performed to maintain and ensure proper functioning of equipment.Pm records were verified to be up to date during the production of this lot.In addition, a device history record review showed no non-conformances associated with this issue during the production of this batch.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd insyte¿ autoguard¿ shielded iv catheter needle would not retract.This complaint was created to capture the 2nd of 2 related incidents.The following information was provided by the initial reporter: "we have had 2 instances where the above needle did not retract.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15643229
Report Number1710034-2022-00611
Device Sequence Number20450653
Product Code FOZ
UDI-Device Identifier00382903814374
UDI-Public(01)00382903814374
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K201075
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/31/2024
Device Catalogue Number381437
Device Lot Number1300906
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer11/29/2022
01/09/2023
Initial Report FDA Received Date10/20/2022
Supplement Report FDA Received Date12/07/2022
01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-