| Model Number |
M2736A |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the ctg (cardiotocograph) machine spontaneously turned itself back on and then recorded a trace which was tachycardic at 180bpm for over 90mins.The device was not in use at the time of the event as the monitor was not hooked up to the patient, no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Reporting address state (b)(6).Reporter institution phone number (b)(6).Reporter phone number (b)(6).A follow up report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported the ctg (cardiotocograph) spontaneously turned itself back on and then recorded a trace which was tachycardic at 180bpm for over 90mins.The device was not in use at the time of the event as the device was not hooked up to the patient, no patient impact.A philips field service engineer (fse) went onsite to evaluate the device in question.Based on the information available and the testing conducted, the cause of the reported problem was a defective transducer.The reported problem was confirmed.The customer was provided with a replacement transducer to resolve the issue.
|
| |
|