MAUDE Adverse Event Report: PERFUSION SYSTEMS AFFINITY PIXIE PEDIATRIC HOLLOW FIBER OXYGENATOR; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS AFFINITY PIXIE PEDIATRIC HOLLOW FIBER OXYGENATOR; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BBP211
Medical Device Problem Codes Gas Output Problem (1266); Increase in Pressure (1491); Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Medtronic received information that during use of a pixie oxygenator, it was reported that the clinician noted there was a difference between the 'dialed in' pump flow on the quantum heart and lung machine (hlm) controller and that on the arterial flow probe which was located between the oxygenator and patient.The clinician was concerned there was something wrong with blood flow delivery and increased the occlusion on the roller pump.They felt the tubing exiting the roller pump (between the pump and oxygenator) and described it as 'rock solid' to the touch indicating high pressures.Pre-oxygenator pressures are not measured.On increasing fio2 as per protocol for cyanotic children, the oxygenation did not increase as expected and the cerebral oximetry started to decrease from 80 to 40% (massimo nirs).Increasing the pump flow did not result in increased flow on the flow probe closest to the patient and manually palpated pressure remained great in the tubing.The patient was ventilated by the anesthetist, bypass was terminated and the oxygenator was changed out.Bypass recommenced, the two flows matched, pressure in the tubing was normal and oxygenation was within the expected range.There was no patient impact associated with this event.Prior to cardiopulmonary bypass (cpb), the majority of the pump prime was removed by retrograde autologous priming (rap) and on initiation of cpb 1000iu heparin was given.As per the facilities local policy, the child was cyanotic pre-cpb so cpb was commenced at a low fio2 to match the child (21%) so as expected venous saturations were low. the clinician questions a transmembrane pressure event or oxygenator blockage.A roller pump was used.No shunts were open other than sample manifold which would not account for the difference in pump flows.Act was measured using a actalyte 'grey tube'.The first act on cpb was 380s and 1000iu was given.Significant haematuria was seen and this cleared later in the bypass.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Device evaluation summary: visual inspection showed evidence of clotting/fibrin.Pressure integrity testing showed no internal or ex ternal leaks when run at 1 lpm with 23 psi, (1189 mmhg) of back pressure for 10 minutes.The device was sent to the blood lab for performance testing.Testing was conducted at a 1:1 ratio (2lpm blood gas flows) the results were as follows; 226 mm/hg delta pblood.The reason for return was undetermined.Conclusion: the complaint was not confirmed for the pixie oxygenator's pressure drop across the device.The issue was not verified via analysis of the returned device.Review of this devices¿ device history record found no abnormalities or non-conforming material reports (ncmrs) initiated during manufacturing that would cause or contribute to the reported event.This device passed all testing and inspections during manufacturing.It is suspected that the high pressure experienced by the customer may be related to a cold induced protein build up or cryoprecipitate which caused initial flow restriction in the device, or the patient or clinical condition at time of bypass or related to heparin potency.The root cause is undetermined.There were no patient/clinical safety issues reported.Trends for issues with this device are reviewed at quarterly quality meetings.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY PIXIE PEDIATRIC HOLLOW FIBER OXYGENATOR
Common Device Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15648383
Report Number2184009-2022-00278
Device Sequence Number20342484
Product Code DTZ
UDI-Device Identifier20613994617310
UDI-Public20613994617310
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K183511
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model NumberBBP211
Device Catalogue NumberBBP211
Device Lot Number223564729
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer10/21/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age4 YR
Patient SexMale
Patient Weight19 KG
-
-