• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT CAPTIVIA - FF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT CAPTIVIA - FF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number VAMF4040C200TU
Medical Device Problem Codes Leak/Splash (1354); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Rupture (2208); Insufficient Information (4580)
Date of Event 10/11/2022
Type of Reportable Event Death
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Valiant captivia stent grafts were implanted during the endovascular treatment of a 175mm thoracic dissection.It was reported approximately 5 years post the index procedure, the patient was transported to the icu after blood pressure started to drop.A non-contrasted ct was performed which did not demonstrate any internal problem but the patents condition continued to worsen.The patient was transferred to the operating room where the chest and abdomen were opened.The physician noticed that the thoracic abdominal aorta had ruptured.The physician decided to extend the existing valiant graft distally to the level of the celiac and complete an open surgical repair on another date.It was noted that the patient's thoracic anatomy was extremely tortuous with 3 acute bends and the physician had a very difficult time getting the valiant grafts to track in the aorta.During the surgery, the physician noticed a large fenestration from the true lumen to the false.The patient's false lumen was very large and aneurysmal (size not known).The physician completed ballooning of the graft at the level of the celiac.It was reported the patient expired on the (b)(6) 2022.Per the physician the cause of the rupture and death are undetermined.
 
Additional Manufacturer Narrative
B5; additional information received ; it was confirmed a reliant balloon was used during the procedure.It was confirmed there was no hole in the existing valiant graft.Per the physician the entire aorta had degenerated with the false lumen expanding since implant.There were multiple fenestrations throughout the aorta.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA - FF
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15649063
Report Number9612164-2022-03879
Device Sequence Number19428422
Product Code MIH
UDI-Device Identifier00643169346543
UDI-Public00643169346543
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/08/2018
Device Model NumberVAMF4040C200TU
Device Catalogue NumberVAMF4040C200TU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2022
Supplement Date Received by Manufacturer10/31/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date11/23/2022
Date Device Manufactured04/08/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."
Outcome Attributed to Adverse Event Required Intervention; Death;
Patient Age69 YR
Patient SexFemale
-
-