MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CONQUEST; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CONQUEST; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQ7586
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly difficult to be removed from the sheath after deflation.It was further reported that the balloon and the sheath was removed together as a one unit.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Expiration date: 08/2024.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon was allegedly difficult to be removed from the sheath after deflation.It was further reported that the balloon and the sheath was removed together as a one unit.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one conquest pta dilatation catheter loaded into an unknown sheath was received for evaluation.During visual evaluation, no kinks were noted to the catheter shaft and no anomalies to the luers, bifurcate or glue fillets.During functional evaluation, the sheath was pulled proximally to remove the balloon but was unsuccessful.The distal end of the sheath was noted to be buckled.No further testing could be performed due to the condition of the sample.Therefore, the investigation is confirmed for the reported sheath removal difficulty as the catheter was received loaded into an unknown sheath and attempts to remove it were unsuccessful.A definitive root cause for the alleged sheath removal difficulty could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 08/2024), g3 h11: h6 (method, result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15649129
Report Number2020394-2022-00826
Device Sequence Number2530505
Product Code DQY
UDI-Device Identifier00801741063121
UDI-Public(01)00801741063121
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQ7586
Device Catalogue NumberCQ7586
Device Lot NumberREFW4451
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer12/02/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-