Model Number DFT415
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that, during intraocular lens (iol) implant procedure, the surgeon found that, there is fracture in the optic, for which he decided to explant the lens and implant another with the same characteristics.Additional information was requested.
 
Additional Manufacturer Narrative
The lens was returned positioned incorrectly in the lens case.Solution was dried on the lens.The optic was cut from edge towards center, typical of insertion and removal.The optic had a crack near the cut damaged area.Complaint history and product history records were reviewed and the documentation indicated the product met release criteria.The file indicated the use of a qualified cartridge and viscoelastic.The handpiece indicated is not qualified for this lens/cartridge combination.The reported optic damage was observed.The root cause for the reported complaint appears to be related to a failure to follow the instructions for use (ifu).The handpiece indicated is not qualified for this lens/cartridge combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Company recommends using the qualified company intraocular lens (iol) delivery system.Per the device ifu: the qualified cartridge/iol combinations have been validated at an ambient temperature of 18 °c using the driving console setting (1.7 mm/sec, 3 seconds, and 3.0 mm/sec for initial velocity, pause and final velocity respectively).Using a higher velocity and shorter pause at lower temperatures, especially with high diopter lenses, could induce damage to iol and/or iol cartridge, affecting successful iol implantation.There are no other complaints in this lot.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY TORIC EXTENDED VISION IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15649187
Report Number1119421-2022-02241
Device Sequence Number14943735
Product Code POE
UDI-Device Identifier00380652422123
UDI-Public00380652422123
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/12/2026
Device Model NumberDFT415
Device Lot Number15218458
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer11/08/2022
Initial Report FDA Received Date10/21/2022
Supplement Report FDA Received Date12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC VISCOELASTIC; INFINITI VISION SYSTEM, ACCESSORY, HANDPIECE; INTREPID AUTOSERT IOL; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; UNSPECIFIED OPHTHALMIC VISCOSURGICAL DEVICE
Patient SexFemale
-
-